Record the calibrate tool's accepted run on image 20260829220329
CAMPAIGN-LOG: aa-offset-calibrate with its 3 s at 8 Hz sampling reads 16.2 counts with a spread of 4.5 over six edges, the value already applied; the compensation's whole path stands on the bench. BRINGUP item 21 keeps only a second machine's value and the lit-tube toggling. No catalog consequence: documentation.
This commit is contained in:
+2
-3
@@ -1477,9 +1477,8 @@ Open items only. Anything closed is in `CAMPAIGN-LOG.md`.
|
|||||||
assist runs unless `cool_aa_offset_counts` carries the machine's value
|
assist runs unless `cool_aa_offset_counts` carries the machine's value
|
||||||
(the `aa-offset-calibrate` diagnostic measures it, the panel's Apply
|
(the `aa-offset-calibrate` diagnostic measures it, the panel's Apply
|
||||||
writes it; zero is the factory's uncorrected reading; the bench
|
writes it; zero is the factory's uncorrected reading; the bench
|
||||||
machine carries 16). Owed: the calibrate tool's recommendation
|
machine carries 16). Owed: a second machine's value when one is on
|
||||||
accepted on an image (its first sampling refused its own result), a
|
the bench; and the mid-run
|
||||||
second machine's value when one is on the bench; and the mid-run
|
|
||||||
toggling between two levels (0.6 to 1.1 C, both sensors together),
|
toggling between two levels (0.6 to 1.1 C, both sensors together),
|
||||||
which comes only with the tube lit: not with the fans alone, not under
|
which comes only with the tube lit: not with the fans alone, not under
|
||||||
motion, not in an armed dark window. The HV supply's input current on
|
motion, not in an armed dark window. The HV supply's input current on
|
||||||
|
|||||||
@@ -4524,6 +4524,17 @@ left `cool_flow_check_s` at 0 (the script's restore posted an empty value
|
|||||||
and got a 400; fixed); the setting was put back to 50 and the engine
|
and got a 400; fixed); the setting was put back to 50 and the engine
|
||||||
re-read it at the next session.
|
re-read it at the next session.
|
||||||
|
|
||||||
|
## 2026-08-29: the calibrate tool's recommendation, on image 20260829220329
|
||||||
|
|
||||||
|
forgectrl 42cdb71 (each edge read over 3 s at 8 Hz a side), image
|
||||||
|
20260829220329 flashed by the operator. `aa-offset-calibrate` from the
|
||||||
|
panel's API, dark: six edges 15.5/16.1, -14.3/-14.2, 16.2/17.3,
|
||||||
|
-13.8/-14.8, 17.8/17.4, -18.1/-18.3 counts (down/up), mean 16.2, spread
|
||||||
|
4.5, recommendation 16.2, which is the value already applied. The
|
||||||
|
compensation's whole path stands on the bench: the tool measures the
|
||||||
|
machine's number, Apply writes it, and both coolant readings hold within
|
||||||
|
0.1 C when the run airflow comes on.
|
||||||
|
|
||||||