ScottW514 89731e654b Record the bench proof on image 20260829190323 and the warm-loop flow check
CAMPAIGN-LOG: the three flowload t1 runs on the new image (judged rise
11.3, 11.1 and 11.8 C with the tube lit through most of each window, the
laser's share on the line), and Test 3, the flow check from a warm loop
(23.6 to 24.9 C baselines, 0 of 6 wrong, the 14.4 C limit 2.25 C above the
flow band and 3.67 C below the no-flow band).

flow_warm_validate.py takes the warm target and the warm-up budget as
arguments (defaults 28 C, 20 min), and the bench page passes them; the
warm-up still judges on the upstream sensor beside the heater, which is
why this run's baselines stopped near 25 C (BRINGUP item 1).

BRINGUP: item 1's open half is the loop above 25 C; item 22 holds what
remains of the flow check under load.

No catalog consequence: a bench tool's arguments and documentation; no
runtime behavior of the release image changes.