Remove history narrative and audit references from docs and tooling

temp_calibrate now cross-checks the factory B-equation instead of the retired linear guess.
This commit is contained in:
ScottW514
2026-08-03 11:54:27 -04:00
parent f92147093e
commit f396dbde34
14 changed files with 76 additions and 79 deletions
+2 -2
View File
@@ -1,9 +1,9 @@
#!/usr/bin/env python3
"""Phase 3.2 software check: laser PWM carrier frequency from PWM2 registers.
"""Software check: laser PWM carrier frequency from PWM2 registers.
Expected with the fsl,extra-prescale=<13> fix: PWMCR prescaler divider = 13,
PWMPR ~125 (127 counts - 2), effective carrier = perclk / (13 * 127) ~= 40 kHz.
Without the fix: divider 1 -> ~520 kHz (audit M8). Safe: read-only register
Without the fix: divider 1 -> ~520 kHz. Safe: read-only register
inspection; the laser PWM output feeds the PSU power input, firing stays gated
by the hardware chain. The definitive gate remains the scope on LASER_PWM.
"""