Add the flowload drill and record the flow check under laser load

The flowload drill in scripts/bench/live_fire_drills.py runs the two tests
of the flow-check plan: t1 fires two CW fills on the press with the check
at its defaults, t2 fires one fill of a chosen length with the check off,
and fit reads rise against dose over the t2 records with the ADC offset
steps masked. The sampler adds the heater output, /cool/status is polled
at 1 Hz with the fan gates, and every controller reply is kept. The job
is fed against the RX buffer's free count, M5 is acknowledged before a
run, a run is refused while the window is armed, and M2 is acknowledged
and the window's close is waited for.

BRINGUP gets item 20: the arm at the first laser-on is skipped while the
driver's spindle-state record reads on, the record is not cleared on
disarm, and the serial layer drops bytes on a full RX ring, so a job with
a lost M5 lets the next job run unarmed. CAMPAIGN-LOG records the Test 1
and Test 2 runs and their numbers.

No catalog consequence: a bench drill on the dev image and documentation;
no runtime behavior of the release image changes.
This commit is contained in:
ScottW514
2026-08-29 13:54:41 -04:00
parent f84b9df040
commit b5de7cda9d
4 changed files with 940 additions and 9 deletions
+88
View File
@@ -4138,6 +4138,94 @@ are not owed. The item is removed from BRINGUP, and the items after it are
renumbered: 17 to 20 are now 16 to 19. `GFSINK_LEAD_MS` (default 10) and
the per-run margin report stay as shipped.
## 2026-08-29: the flow check under laser load, Tests 1 and 2
The 2026-08-25 flow-check trip (heater rise 15.1 C against the 14.4 C limit,
dT 9.4, while the first `dpatch` patch fired CW through the check window) was
run down with a new drill, `flowload` in `scripts/bench/live_fire_drills.py`.
`flowload t1` puts the three check keys at their defaults for the run and
fires two 30 x 4 mm CW fills (F1500, 0.3 mm pitch, about 35 s lit) on the
press with no dark dwell; `flowload t2 <secs> [pct]` writes
`cool_flow_check_s = 0` for the run and fires one fill sized to the lit
seconds at CW or at a density level; `flowload fit` fits rise against dose
over the t2 records. The sampler is the `dpatch` one plus
`thermal/heater_pwm`, `/cool/status` is polled at 1 Hz with the fan gates,
and every controller reply is kept. Every key the drill writes goes back to
what stood before when the run ends. The pump was never commanded off.
Records: `bench-data/flowload_t1_20260829-*.json` and
`bench-data/flowload_t2_*_20260829-*.json` in the tree.
**Test 1, three runs.** The check starts at the session open (the heater
comes on about one second after the M3, not at the press), so the press must
come at once for the fire to overlap the window. Runs 1 and 2 ended their
windows early (the drill closed the session on M2 before the 50 s were up;
fixed: the drill now holds the window open until the heater trace ends).
Run 3 ran the full window with the tube lit for 70 % of it, and the engine
read `coolant flow verified (heater rise 14.1 C, dT 9.5 C)`: 0.3 C from a
SUSPECT, where the same loop reads 11.7 to 12.1 C dark. The trip is
reproduced in kind, and it is not flow. Two things stack:
- **A common-mode ADC offset while the run airflow profile is on.** One
sample after the session opens (fans to run duty) both coolant sensors
drop 1.0 to 1.9 C together; they step back up when the fans return to
idle; in between the readings toggle between two levels 0.6 to 1.1 C apart,
both sensors in lockstep, up to 22 times in a run, with every fan steady
at speed. The one session in which the air-assist fan never left idle
showed no step and no toggling, the only pointer to a source so far. The
engine captures `flow_base_down` from one sample at its first tick after
the heater starts, inside that offset, so every rise carries about +1.1
to +1.4 C from the offset and about +-0.5 C of single-sample scatter, dark
or lit.
- **The tube's heat at the sensors.** Test 2 below: about 1.5 C inside a
fully lit 50 s CW window.
Dark 11.7 plus the tube's 1.5 plus one low base sample reaches 14.1; the
15.1 trip is the tail of the same distribution.
**Test 2, the tube's signature, check off.** CW bursts of 20.8, 40.8, 47.9
and 59.0 s and one 59.4 s burst at 45 % density (S450). With the ADC offset
steps masked (a step is both sensors' half-second mean levels changing by
0.45 C or more the same way, agreeing within 0.4 C, subtracted from all
later samples), the downstream rise at burst end against the `hv_current`
integral is linear through the origin: **k = 3.06e-5 C per raw-second**
(r2 0.981, intercept 0.04 C), 0.030 C per lit second at hv 971, so **a fully
lit 50 s CW check window adds 1.49 C** against the 1.6 C margin; the rise
50 s after fire start read 1.43 to 1.49 C on every burst long enough. The
lag from first emission to the first sensor response is 10 to 20 s, a
smooth ramp on both sensors together, never a step at fire start. At 45 %
density the same window adds 0.46 C, and the heat per raw-second is 0.77 of
CW: the current integral overstates density heat, so a tracer needs one k
per power model.
**Events on the way.** One `t2 40` run held at +7 s on the airflow gate
(`air_assist 1895 under the 6000 floor for 3 s`): the air-assist fan never
left its idle reading inside the 15 s grace, the first air-assist trip on
record; it spun up normally in every other session. The first `t2 60` run
was written to the controller as one 93-line block, about 1270 bytes
against the 1023-byte RX ring, and the serial layer drops bytes on a full
ring, so the fill ran 46 s instead of 60 and the job's M5 and M2 were lost:
the window stayed open (engine phase `run`, `armed` true) until the drill
exited and dropped the connection. The next `t2 60` then ran its whole fill
with no arm, no button wait, no run report and no airflow: the driver's
spindle-state record was still on from the lost M5, and the arm at the
first laser-on is skipped when that record reads on. Fire stayed suppressed
at the stream (no HV, no emission, thermopile flat), so no energy left the
tube, but the head ran a full job without the operator's press. This is
BRINGUP "Next work" item 20 (arm on `state.on && !laser_ok`, clear the
spindle state in `gflaser_disarm`, consume the RX overflow flag), a fix
owed before the next image. The drill now feeds the job against the `Bf:`
free-character count, sends and acknowledges M5 before every run, refuses
to start while `/cool/status` shows the window armed, acknowledges M2 and
waits for the window to close, and prints the controller's replies: the
last three runs show `press the button to start the laser job`,
`laser armed (density)` on the press, and `Pgm End` with
`laser disarmed - latch locked` on the M2.
**Also seen.** `laser power-good degraded during the armed window` is
warned by the engine at every session open, a separate item. The check
window's own baseline and the tube term are the two candidates the fix
chooses between; nothing is built yet.