bench+docs: flow suspicion drills, coolant-flow triage record

flow_confirm_drill.py walks the driver's suspicion/confirmation state
machine through every verdict with real pump-off transients in one M8
session; flow_escalate_drill.py exercises the starved-re-check
escalation against a short GFCOOL_CONFIRM_MAX_S. BRINGUP records the
triage resolution (the 2026-08-03 faults were a real transient
stagnation, probable pump airlock - the check was right), the slug/
circulation measurements, and the new check semantics.
This commit is contained in:
ScottW514
2026-08-08 10:59:47 -04:00
parent 548d25a4ff
commit af3a693c0b
3 changed files with 403 additions and 11 deletions
+85
View File
@@ -0,0 +1,85 @@
#!/usr/bin/env python3
"""Bench drill for the starved-re-check escalation path.
Runs ON THE BOARD against a controller started with a short
confirmation budget (GFCOOL_CONFIRM_MAX_S=45; see the runbook for the
manual start line). With the pump off, the job-start check reads
over-limit -> SUSPECT; the cooked stagnant loop then cannot pass the
settle gate within 45 s, so the budget expires and the driver must
escalate: "COOLANT FLOW FAULT: no clean re-check within 45 s".
Prints PASS/FAIL and leaves the machine idle: M9 sent, pump on,
heater off.
"""
import select
import socket
import time
T0 = time.monotonic()
def el():
return time.monotonic() - T0
def log(msg):
print('%7.1f %s' % (el(), msg), flush=True)
def pump(on):
with open('/sys/glowforge/thermal/water_pump_on', 'w') as f:
f.write('1' if on else '0')
log('** pump -> %s' % ('ON' if on else 'OFF'))
s = socket.create_connection(('127.0.0.1', 23), timeout=5)
s.setblocking(False)
buf = b''
def wait_msg(substrs, deadline):
global buf
while el() < deadline:
r, _, _ = select.select([s], [], [], 0.2)
if not r:
continue
data = s.recv(4096)
if data:
buf += data
while b'\n' in buf:
line, buf = buf.split(b'\n', 1)
line = line.strip().decode('ascii', 'replace')
if line.startswith('[MSG'):
log('<< ' + line)
for m in substrs:
if m in line:
return line
return None
ok = True
try:
time.sleep(0.5)
pump(False)
log('--- M8 with the pump off')
s.sendall(b'M8\n')
if wait_msg(['FLOW SUSPECT'], el() + 180) is None:
log('!! no SUSPECT within 180 s')
ok = False
elif wait_msg(['no clean re-check within'], el() + 120) is None:
log('!! no escalation FAULT within 120 s of the suspect')
ok = False
finally:
try:
s.sendall(b'M9\n')
time.sleep(1.0)
except OSError:
pass
s.close()
pump(True)
with open('/sys/glowforge/thermal/heater_pwm', 'w') as f:
f.write('0')
log('--- M9 sent, pump on, heater off')
print('ESCALATION DRILL %s' % ('COMPLETE' if ok else 'FAILED'), flush=True)